UGT ku-pF Apparatus

Item no.: 1435000 (ku-pF MP10) Item no.: 1434000 (ku-pF SP)

Advantages at a glance

  • Testing of up to 10 samples simultaneously
  • Continuous data acquisition witha minimum of monitoring
  • Balancing of the scales between weighings

Please send me more information about UGT ku-pF Apparatus

UGT ku-pF Apparatus

The ku-pF MP10 is a laboratory instrument for the automatic determination of unsaturated hydraulic conductivity ku and the pf curve (water tension curve) in accordance with the evaporation method on up to 10 soil ring samples simultaneously. The ku-pf apparatus works independently of a PC due to the integrated data logger and a microcomputer. A connection must be created to the PC only for maintenance, test start and reading out of the data. The ku-pF apparatus is available as a single measurement station ku-pF SP.

The apparatus is designed for the use of the UGT soil sample rings with 250 cm³ inner volume and 41 cm² cross sectional area. These soil sample rings have bore holes arranged exactly at height intervals of 3 cm for the insertion of the tensiometers. However, they may also be used for obtaining samples separately from the ku-pf apparatus.

The filled soil sample rings are set onto filter plates, saturated and sealed basally in the carrier basket for the test run. The gradient resulting from the condensation on the free ground surface is recorded by two tensiometers which record the corresponding water content and the flow rate by periodical weighing respectively. The Tensio 130 tensiometers with particularly small ceramic cup (Ø = 6,5 mm / L = 20 mm) provide exact measurements with a minimum destruction of the soil. Air bubbles are easy to recognise and eliminate through the acrylic head with bleed screw.

The arrangement of the samples on the star shaped carrier which rotates at the set test interval permits the examination of 10 samples simultaneously. The sample on the weighing position is lifted from the carrier by lifting the scales so that it is standing free. The weighing data and the tensions here are stored on the internal data logger. Automatic measuring produces continuous measurement series with a minimum of monitoring. The scales are automatically balanced between weighings. Drift effects or any foreign bodies on the scales do not therefore cause any interference.

Set consisting of

  • 1 ku-pF apparatus (ku-pF MP10)with internaldata logger
  • 1 laboratory scale
  • 10 carrier baskets
  • 10 tensiometer pairs Tensio 130
  • Drill holes for tensiometer installationincluding 3 augers andguide device
  • 10 soil sample rings V = 250 cm³/ Ø = 41 cm²
  • 20 lids for soil sample rings
  • 10 filter plates
  • 1 software kuLog
  • Perforated sheets
  • Filter paper

 

Parameters

Specifications

Basic device

ku-pF SP

ku-pF MP10

No. of samples / meas. positions

1

1 … 10

Dimensions L x W x H

30 x 45 x 40 cm

100 x 70 x 40 cm

Weight

ca. 13 kg

ca. 40 kg

Material

Housing: Sheet steel, powder-coated

Support arm: Aluminum, powder-coated

Rotary star with support arms: Aluminum, powder-coated

Protection class

IP 40

Operating temperature

+1 … +40 °C

Measurement positions

Data collection cycle time

1 … 4 min

10 … 40 min

Carrier basket

  • For soil sampling rings with an internal volume of 250 cm³, internal diameter Ø 72 mm, height 61 mm
  • Two Tensio 130 units each, supplied with transport/storage tubes
  • 2-channel measurement processor unit
  • Weight: approx. 440 g per carrier basket

Tensio 130 Laboratory tensiometer

  • Tension range: 0 … >85 kPa ±0.3 kPa
  • Ceramic cup: P80, 6.5 × 20 mm, with screw adapter
  • Measurement position: at the weighing position

Sieve plate

One perforated sieve plate per carrier basket (approx. 44 g)

Built-in Data logger

 

Interface

RS-232 interface (115200 baud)

Storage capacity

210 … 840 days

Power supply

 

Supply voltage

230 V

Power consumption

1.2 A

Features

  • Stepper motor drive for automatic sample positioning
  • Weighing table with lifting mechanism
  • Integrated data logger with control functionality for autonomous test operation and measurement data storage

 

Precision Laboratory Scale

Specifications

Dimensions LxWxH

193 x 262.5 x 84.5 mm

Weighing surface

Ø150 mm

Measurement range

0 … 3,200 g

Resolution

0.01 g

Net weight

~3 kg

 

Alu sampling case

Specifications

Dimensions L x W x H

440 x 360 x 145 mm

Weight

6.6 kg

10 Soil sample rings

250 cm³, ∅i 72 mm, 61 mm height

20 lids

Aluminum

Accessories

Polyamide hammer, impact head, spatula

 

Pre-drill set for Tensiometer

Specifications

Drilling device 

  • For mounting the soil sample ring
  • 125 x 95 x 83 mm
  • Aluminum
  • ~564 g

Cork drill with drill core remover

110 mm, Øi 5,5 mm 

Twist drill

106 mm, Ø6,5 mm 

 

Calibration and filling set

Specifications

Hand vacuum pump

-1.0 … +4.0 bar

Manometer 

-1 … 0 bar display, scale division 5 mbar

Vacuum chamber (PVC)

500 ml capacity

Accessories

Tube, adapter, syringe, calibration cable

 